A rare case of reel syndrome: The unexpected permanent pacemaker event

Authors

DOI:

https://doi.org/10.21776/ub.hsj.2025.006.04.16

Keywords:

Reel syndrome, pacemaker complication, early diagnose

Abstract

BACKGROUND: "Reel syndrome" is a rare complication that occurs after implantation, affecting up to 1.7% of patients with large generator pockets. This condition generally manifests 4–6 weeks post-implantation.

CASE ILLUSTRATION: A 64-year-old woman presented unexpectedly with exertional dyspnea and mild fatigue over the past two days. She has a history of pacemaker installation from one month ago. The ECG demonstrated loss of capture, and pacemaker interrogation showed increased capture thresholds with normal impedance. A chest X-ray revealed a ventricular lead coil encircling the generator with lead dislodgment. Reel syndrome is defined by the torsion of the pulse generator along its transverse axis, leading to the coiling of the pacemaker leads around it, which typically protects both the generator and the leads from damage. Symptoms remain asymptomatic until lead dysfunction occurs due to lead malposition. We suggest appropriate diagnostic assessments, such as chest X-rays and ECGs, for prompt identification. The older population has been recognized as a considerable risk factor for this "event."

CONCLUSION: Upon confirmation, reel syndromes require immediate intervention. Effective implantation techniques and information for patients and caregivers are essential in alleviating these issues.

Author Biographies

  • Yos Akbar Irmansyah, Department of Cardiology and Vascular Medicine, Faculty of Medicine, Universitas Sebelas Maret, Surakarta, Indonesia

    Department of Cardiology and Vascular Medicine, Faculty of Medicine, Universitas Sebelas Maret

  • Irnizarifka, Division of Arryhtmia, Cardiac Pacing, and Heart Failure Universitas Sebelas Maret Hospital, Department of Cardiology and Vascular Medicine, Faculty of Medicine, Universitas Sebelas Maret

    Division of Arryhtmia, Cardiac Pacing, and Heart Failure Universitas Sebelas Maret Hospital, Department of Cardiology and Vascular Medicine, Faculty of Medicine, Universitas Sebelas Maret

References

1. Tseng AS, Shipman JN, Lee JZ, Mi L, Amin M, Killu AM, et al. Incidence, patterns, and outcomes after transvenous cardiac device lead macrodislodgment: Insights from a population-based study. Heart Rhythm 2019;16 (1):140-147. doi: 10.1016/j.hrthm.2018.07.026.

2. Adhikari S, Annie F, Amer M. Reel syndrome, Lead Macrodislodgement2023.

3. Hilal S, Chikhi F, Hassan F, Fellat I, Cherti M. Delayed presentation of pacemaker Reel Syndrome and lead damage: Coincidence or consequence? Indian Pacing Electrophysiol J 2022;22 (5):251-253. doi: 10.1016/j.ipej.2022.07.001.

4. Harcombe AA, Newell SA, Ludman PF, Wistow TE, Sharples LD, Schofield PM, et al. Late complications following permanent pacemaker implantation or elective unit replacement. Heart 1998;80 (3):240-244. doi: 10.1136/hrt.80.3.240.

5. Bellinge JW, Petrov GP, Taggu W. Reel syndrome, a diagnostic conundrum: a case report. European Heart Journal - Case Reports 2021;5 (10):1-6. doi: 10.1093/ehjcr/ytab394.

6. Carnero-Varo A, Pérez-Paredes M, Ruiz-Ros A, Gimenes-Cervantes D, Martinez-Corbalan F, Cubero-Lopez T, et al. "Reel Syndrome": A New Form of Twiddler’s Syndrome? Images in Cardiovascular Medicine 1999;100 (1):e45-e46. doi.

7. Alvarez-Acosta L, Romero Garrido R, Farrais-Villalba M, Hernandez Afonso J. Reel syndrome: a rare cause of pacemaker malfunction. BMJ Case Rep 2014;2014 (1):bcr2014204545. doi: 10.1136/bcr-2014-204545.

8. Yuksel S, Demir S, Sahin M. Total extrusion of a normally functioning pacemaker. Tex Heart Inst J 2012;39 (1):156-157. doi.

9. Smit M, Coetzee AR, Lochner A. The Pathophysiology of Myocardial Ischemia and Perioperative Myocardial Infarction. J Cardiothorac Vasc Anesth 2020;34 (9):2501-2512. doi: 10.1053/j.jvca.2019.10.005.

10. Patel MB, Pandya K, Shah AJ, Lojewski E, Castellani MD, Thakur R. Reel syndrome-not a twiddler variant. J Interv Card Electrophysiol 2008;23 (3):243-246. doi: 10.1007/s10840-008-9288-0.

11. Kus G, Cagirci G, Ureyen CM, Bayar N, Arslan S. Unexpected cause of involuntary muscle movements: Reel syndrome. Anatol J Cardiol 2021;25 (7):515-516. doi: 10.5152/AnatolJCardiol.2021.38445.

12. Rocco E, Simeone B, Ciacciarelli M, Petrassi M, Di Rosa R, Versaci F, et al. Reel syndrome: a rare presentation of a rare cause of cardiac resynchronization malfunction. Clin Ter 2022;173 (4):295-296. doi: 10.7417/CT.2022.2434.

13. Gkinos C, Manouras A, Lagoudianakis EE, Papadima A, Tsioufis C, Vavouranakis E, et al. Twiddler's syndrome. Hellenic J Cardiol 2007;48 (5):300-301. doi.

14. Irnizarifka, Arifianto H. The The COmprehensive Registry and rEsearch on Heart Failure (CORE-HF) : 2 years report from single-center Indonesian heart failure clinic registry. ACI (Acta Cardiologia Indonesiana) 2021;7 (2):10. doi: 10.22146/jaci.v7i2.2266.

Downloads

Published

2025-10-31